Tema de trabajo aprobado
ISO/AWI 20411
Surface chemical analysis — Secondary ion mass spectrometry — Correction method for saturated intensity in single ion counting dynamic secondary ion mass spectrometry
Reference number
ISO/AWI 20411
Edition 2
Tema de trabajo aprobado
ISO/AWI 20411
88328
Un grupo de trabajo ha preparado un borrador.
Reemplazará ISO 20411:2018

Resumen

ISO 20411:2018 specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in pulse counting magnetic sector-type secondary ion mass spectrometers or quadrupole secondary ion mass spectrometers. It uses a test based on depth profile analysis of two isotopes in a reference material which has a gradual concentration change between low and high concentration regimes. It also includes a correction method for saturated intensity caused by the dead time of the detector. The correction can increase the intensity range for 95 % linearity so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction equations have been shown to be valid.

ISO 20411:2018 does not apply to time of flight mass spectrometers.

ISO 20411:2018 is only applicable to elements with minor isotopes. It is not applicable if the element is monoisotopic or contains isotopes with equal abundances.

Informaciones generales

  •  : En desarrollo
    : Nuevo proyecto registrado en el programa de trabajo TC/SC [20.00]
  •  : 2
  • ISO/TC 201/SC 6
  • RSS actualizaciones